1) spiral cable wrap (3) cable clips (1) mounting set: hex key 3 mm (for wall channel adapter), 4 mm, 5 mm, 6 mm, M 6 x 40 hex screw Scope of Delivery: Variable Height Arm with Extension Arm (1) Variable Height Arm with Extension Arm (3) end caps (1) aluminium profile for Variable Height Arm (1) aluminium profile for Extension Arm (1) spiral cable wrap (3) cable clips (1) mounting set: hex key 3 mm (for wall channel adapter), 4 mm, 5 mm, 6 mm, M 6 x 40 hex screw |
||
Cable Integration Delivery contains a spiral cable wrap for cables. The spiral cable wrap is to protect the cables against damage and squeezing especially in the pivot zones. Please make sure that the spiral cable wrap goes through both joints. Wrap the spiral cable wrap around the cables. Make sure that the cables are not contorted. The spiral cable wrap organizes up to 3 cables depending on the cable diameter. The spiral cable wrap may not exceed a diameter of 15 mm. Cables may only be integrated when the spiral cable wrap is used. Non-observance may result in damaging the cables. In case you do not wish completely integrated cable management (3) cable clips help manage cables. Fix the cable clips underneath the aluminium profile. |
||||
Cabling - Please follow the steps as described belowMedical Variable Height Arm 1. Remove the end caps (if not removed already). 2. Remove the aluminium profile form the Variable Height Arm (if not removed already). 3. Remove the monitor adaptation by loosing the (4) M 5 x 16 screws. You have the following components now: plastic layer ring, plastic ring, metal ring, (4) M 5 x 16 screws. Please make sure that you re-assemble the monitor adaptation in the reversed order. |
||||
![]() |
||
4. Remove the outer cover form the outer pivot block by loosing the (2) socket head screws M 5 x 20. |
||
![]() |
||
5. Slide the arm into the wall channel or mount it to the European rail or the pole. 6. Remove the inner cover from the inner pivot block by loosing the (2) M 5 x 20 screws. |
||
![]() |
||
7. Disengage the locking lever and bring the arm in position may the inner cover be removed. |
||
![]() |
||
8. Remove the right hex screw M 6 x 12 and replace it with the hex key M 6 x 40. Fasten the screw until the depth gauge without using any force. The M 6 x 40 screw is used to punch the spindle through with a hammer by about 20 mm. Then unscrew the M 6 x 40 screw. |
||
WARNING: Avoid pulling out the spindle completely. |
||||
![]() |
![]() |
|||
9. Lead the spiral cable wrap with the cables from below through the console and from above through the inner pivot block into the support arm profile. Lead the spiral cable warp from the support arm profile downwards through the outer pivot block 10. Re-assemble the support arm in reversed order. |
||
![]() |
![]() |
|||
Medical Variable Height Arm with Extension Arm Proceed in the way as described with the Variable Height Arm step 1 to step 8. Please make sure that you exactly follow the mounting instructions and warnings. Then proceed in the way as described below: 1. Remove the support arm from the wall channel, the European rail plate or the pole mount. Please make sure the support arm is in the lowest position and be careful not to disengage the locking lever by chance. 2. Remove the adapter for the wall channel, the European rail plate or the pole mount from the console and turn it by 180°. Please make sure that you do not forget the (2) locating pins when re-assembling the adapter. |
||
![]() |
||
3. Remove the end caps and the aluminium profile (if not removed already). 4. Depending on the connector type it may be necessary to disassemble the Extension Arm from the Variable Height Arm by loosing the (4) M 5 x 20 screws. You have the following components now: Variable Height Arm, plastic layer ring, extension arm, plastic ring, metal ring, (4) M 5 x 20 screws. |
||
Please make sure that you re-assemble the arm in reversed order. |
||
![]() |
||
5. Put the spiral cable wrap into the support arm profile and turn the console back by 180° to its original position. |
||
![]() |
||
6. Lead the spiral cable wrap with the cables through the Variable Height Arm and follow the instructions described in step 9 Variable Height Arm. 7. Put the spindle back to its position. Follow the instructions described in step 8 Variable Height Arm, but in reversed order. 8. Mount the inner cover to the inner pivot block using (2) M 5 x 20 hex screws. Follow the instructions described in step 6 and 7 Variable Height Arm, but in reversed order. 9. Mount the Extension Arm to the Variable Height Arm using the components in reversed order. 10. Mount the adapter for the wall channel, the European rail adapter or the pole mount to the console. Please make sure that you do not forget the (2) locating pins. 11. Slide the support arm into the wall channel or mount it to the European rail or pole. Disengage the locking lever and put the arm in the highest position. 12. Mount the outer cover to the outer pivot block using (2) M 5 x 20 hex screws. Follow the instructions described in step 4 Variable Height Arm. 13. Mount the monitor adaptation to the support arm. Follow the instructions described in step 3 Variable Height Arm, but in reversed order. |
||||
Electric Safety To comply with the regulations for electric safety in medical environments (International Standards IEC) an additional equipotential bonding connector is provided. In order to eliminate any difference in potential between touchable parts, ground and other conductive parts and to protect people from electric shock make sure that the support arm is connected to the potential bonding. Furthermore, consider the national electric safety regulations. Mounting the Monitor 1. Ensure that the locking lever is engaged with the arm in the highest vertical position. Do not disengage the locking lever unless an instrument is mounted on the end of the support arm. 2. Slide the support arm into the ErgonoFlex wall channel or mount it to the European rail or pole and tighten the set screws. 3. Only after set screws have been tightened can medical devices be mounted. For most devices a specific adapter plate provided by ErgonoFlex med is needed. Please refer to assembly instructions provided with the adapter plate. 4. Once the adapter plate has been attached to the device, secure it to the mounting plate of the support arm. Slide the device into the mounting plate until plunger locates in the clearance hole. Tighten the nylon screws in the mounting plate. With VESA adaptations the flat panel is attached directly to the VESA mounting plate at the front of the support arm. Please use the M 4 screws with the suitable length. If needed nylon spacers can be used. Tilt, Swivel and Pivot Adjustments To adjust the overall tilt tension tighten or loosen the (2) hex screws with the hex key 4 mm provided. To adjust the swivel and/ or pivot tension remove the end caps and tighten or loosen the (4) head screws. |
||||
![]() |
||
WARNING For devices mounted from the bottom the tilt angle is limited to 20°. Never attempt to loosen the hex screws and the locking lever when the monitor is mounted to the arm. Always remove the monitor first. Cleaning ErgonoFlex support arms may be cleaned with mild, non-abrasive solutions commonly used in hospital environment. Do not use steel wool or other abrasive material. |
||||
WARNING Never attempt to remove the support arm from the ErgonoFlex wall channel when loaded with the device. When removing the instrument, please verify that the locking lever is engaged with the arm in the highest vertical position.
Wall Mount Overview |
||||
![]() |
||||
![]() |
||||
![]() |
||||
![]() |
||||
![]() |
||||
![]() |
||||
![]() |
||
|
||||||||||||
ErgonoFlex® /Createch SPRL
Vorstsesteenweg 78
1601 Ruisbroek - Belgium
Tel: +32 2 538 33 38
Fax: +32 2 537 27 42
ErgonoFlex® is a registered trademark of Createch SPRL
Advanced Workplace Wellness
ErgonoFlex Medical Monitor Arms Assembly and Technical Instructions
Medical Monitor & Laptop Mount |